Home  /  Analytes / Retatrutide

Retatrutide

What identity means for this molecule in numbers, how we measure its purity, and whether that method is validated yet. CAS 2381089-83-2 Not yet checked against the CAS registry.

Identity by mass spectrometry

Identity by LC–MS means every charge state we record lands within 10 parts per million of where the formula says it must. That establishes the formula of the main component; the limits page says what it does not establish. The windows below are that criterion in daltons, at each m/z.

Molecular formula

C221H342N46O68

Monoisotopic mass

4728.4718 Da, free peptide

Average mass

4731.35 Da, free peptide

Expected charge states of Retatrutide in the scan range, with the 10 ppm identity window at each and the spacing of its isotope peaks.
Statem/z 10 ppm, Da10 ppm, mDa Isotope spacing, m/z
[M+3H]3+1577.1645±0.0158±15.80.334
[M+4H]4+1183.1252±0.0118±11.80.251
[M+5H]5+946.7016±0.0095±9.50.201
[M+6H]6+789.0859±0.0079±7.90.167
[M+7H]7+676.5032±0.0068±6.80.143

Scan range 300–2200. The bold state is the one expected to be most abundant. We expect the envelope to centre on the smaller of M/1000 + 1 and the larger of the basic-site count and M/1000, M in daltons; here that is M/1000 = 4.7. It has 2 basic sites, and a peptide this size also carries protons on backbone amides and on Gln and Asn side chains. A molecule this size spreads across several states, and every state we record, at least the two most abundant, has to agree.

The last column is the spacing between neighbouring isotope peaks of each ion: the mean step one heavy isotope adds to this composition, 1.0030 Da, divided by the charge. The higher the charge, the closer the peaks; at [M+7H]7+ they sit 0.143 m/z apart, one part in 4 700 of its m/z. Resolving power is m/z over the smallest separation an instrument shows, so the method’s floor of 10 000 at m/z 1000 shows peaks 0.1 m/z apart there, against 0.20 m/z between the isotope peaks of a 5+ ion at that m/z: that is why the 10 ppm criterion is applied only on an instrument of that resolving power or better, and why a unit-resolution instrument, which shows peaks about 1 m/z apart, cannot pick the monoisotopic peak out of any cluster past 1+.

Before you send. The formula reproduces from the sequence only with one AEEA unit in the linker (γGlu-AEEA-C20 diacid on Lys17); with two it would be C227H353N47O71.

Expected mass spectrum, LC–MS, ESI+ · RetatrutideCharge-state envelope for a peptide of monoisotopic mass 4728.47 daltons: the 5+ ion is the base peak at m/z 946.70, with 3+, 4+, 6+, 7+ either side. Each state is drawn as an isotope cluster. 1577.16 [M+3H]3+ 1183.13 [M+4H]4+ 946.70 [M+5H]5+ 789.09 [M+6H]6+ 676.50 [M+7H]7+ 600 800 1000 1200 1400 1600 0 50 100 m/z rel. % LC–MS, ESI+ · Retatrutide expected · monoisotopic mass 4728.47 Da
Expected envelope · computed The charge states a 4.7 kDa peptide should present in ESI+, placed by arithmetic from the formula. Relative intensities are a model. Nothing here is a measurement; the measured spectrum appears on the report.

Method and validation status

Purity is area percent by RP-HPLC–UV 214 nm. Identity is LC–MS (ESI+). The matrix we accept is lyophilised powder.

Status

Validation
Not yet validated
Retention time
not yet measured on this analyte
Limit of quantitation
not yet measured
Reporting threshold
0.05% area; no limit of quantitation is claimed yet, so a smaller peak may be present and not seen

Not yet validated means the method is written and has not been run on this analyte. The retention time and the limit of quantitation are listed once measured, with the date of the validation that produced them. The scope table shows the status of every analyte.

What the formula tells us to watch for

  • Deamidation (+0.984 Da) is sought on any sequence with asparagine or glutamine. The sequence has them; where it applies it is confirmed on the chromatogram as a closely eluting peak, before or after the parent, since at charge z the mass shift is 0.984/z m/z, which at 4+ and above overlaps the M+1 isotope.
  • Truncation products - the sequence short by one or more residues - fall at characteristic mass offsets and are looked for at their expected m/z where the sequence is on file; the certificate lists the impurity masses that were read and whether each was assigned.

Order a test of Retatrutide

Identity and purity together, $185, in 10 business days or the invoice is cancelled, or the payment refunded, and you still get the results. One sealed container as received.

What you get back

A certificate with the chromatogram, every peak, the three injections, the measured spectrum against the windows above, what could not be resolved, a full signature, a hash and a QR that verifies without us. How to read one.

Identity + Purity, $18510 business days or they’re free
Order